How little can still be measured? (calibration line, decision limit, limit of detection and quantification)
In brief
In the laboratory x is the known amount (concentration) and y the measured signal. The calibration line connects the two. Three limits follow from it: the decision limit (Nachweisgrenze: smallest amount that can be told from the zero value), the limit of detection (Erfassungsgrenze: smallest amount that is detected reliably) and the limit of quantification (smallest amount that can be measured with acceptable uncertainty).
When
- At least four calibration points; DIN 32645 requires at least six at equal distances. With fewer points the limits are very uncertain.
- The points must really lie on a straight line, their spread should be similar over the whole range, and the range should begin close to zero.
How it works in Statable
- Run the card "Are two things related?" with the amount as x and the signal as y; the scatter plot appears below.
- Open the section "Calibration line (limit of detection and quantification)". Choose the "Error probability": 1% (default, mostly used in the DIN) or 5%.
Reading the result
- The line is given as an equation. The table "Limits of the procedure" gives decision limit, limit of detection (twice the decision limit, an approximation) and limit of quantification (smallest amount with at most 33% relative uncertainty), plus the method standard deviation (residual spread of the line, converted to the unit of x) and the method coefficient of variation (this divided by the mean of x).
- If the line does not rise, the signal does not depend on the amount and there are no limits. If the line scatters too strongly, the limit of quantification cannot be calculated.
What is often misunderstood
The limits hold only within the assumptions (straight line, similar spread over the whole range). The scatter plot shows whether that is true. Statable assumes a single measurement of the sample.
Reporting the result
Report: the number and range of the calibration points, the equation of the line, the method and the error probability, then decision limit, limit of detection and limit of quantification with unit, plus the standard deviation of the procedure and its coefficient of variation. The calibration-line method is standardized in DIN 32645 ("decision limit, detection limit and determination limit under repeatability conditions"). Also: anyone reporting under ICH Q2(R1) calculates with 3.3 · σ / slope and 10 · σ / slope; the numbers are then not the same, and the method must be named.
The calibration line (n = … calibration points, range … to … [unit]) is y = … + … · x. By the DIN 32645 procedure (error probability α = …) the decision limit was … [unit], the detection limit … and the quantification limit … (k = 3); the standard deviation of the procedure was … and its coefficient of variation … %.
For specialists
Line y = a + b x (least squares), s = s_yx, Qxx = Σ(x - x̄)², n points, m = 1. Decision limit x_DL = (s / b) · t(f; 1 - α) · √(1/m + 1/n + x̄² / Qxx), one-sided, f = n - 2. Limit of detection x_LOD = 2 x_DL. Limit of quantification x_LOQ = k (s / b) · t(f; 1 - α/2) · √(1/m + 1/n + (x_LOQ - x̄)² / Qxx) with k = 3, quadratic in x_LOQ and solved in closed form. Reference values: R chemCal (lod, loq).
Sources
DIN 32645:2008-11. Chemical analysis - Decision limit, detection limit and determination limit under repeatability conditions. Beuth. Currie, L. A. (1968). Limits for qualitative detection and quantitative determination. Analytical Chemistry, 40, 586-593.